ORIGINAL ARTICLE
Year : 2021  |  Volume : 33  |  Issue : 4  |  Page : 422-430

Corneal haze and densitometry in keratoconus after collagen cross-linking by three different protocols


Department of Ophthalmology, Mansoura Ophthalmic Center, Faculty of Medicine, Mansoura University, Mansoura, Egypt

Date of Submission02-May-2021
Date of Decision06-Jun-2021
Date of Acceptance06-Jun-2021
Date of Web Publication06-Jan-2022

Correspondence Address:
Amani E Badawi
Department of Ophthalmology, Mansoura Ophthalmic Center, Faculty of Medicine, Mansoura University, 60 Elgomhoria Street, Mansoura
Egypt
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/joco.joco_145_21

Rights and Permissions
  Abstract 


Purpose: To evaluate postoperative corneal haze and corneal densitometry following three different corneal cross-linking (CXL) protocols; standard, accelerated, and trans-epithelial (TE).
Methods: The study recruited 104 eyes (53 patients) with progressive keratoconus divided into three groups: Group I were subjected to standard CXL, Group II to TE-CXL, and Group III to accelerated CXL (A-CXL) (10 mW/cm2 for 9 min). Subjective and objective corneal haze measures were evaluated before and 3, 6, and 12 months post-CXL using slit-lamp biomicroscopy and Pentacam Sheimpflug camera.
Results: There was a significant difference in corneal densitometry between the three groups at 3 and 6 months post-CXL (P < 0.0001). By the 12th month, a significant statistical difference was observed only in zones (0–2 mm) and (2–6 mm) in both the anterior and the central layers. In Group I, the densitometry value of the preoperative anterior stromal layer (anterior 120 μm) was 19.42 ± 1.81. Then, it peaked at 23.12 ± 1.21 at 3 months (P < 0.0001), reached 19.82 ± 1.19 at 6 months (P = 0.007), and decreased to 19.33 ± 3.23 (P > 0.05) at 12 months. In Group II, the preoperative densitometry value of the anterior layer was 19.41 ± 1.21, peaked at 19.72 ± 1.12 at 3 months (P = 0.02), reached 19.04 ± 1.18 at 6 months (P = 0.052), and increased to 19.13 ± 1.37 at 12 months (P = 0.84). In Group III, the preoperative densitometry value of the anterior stromal layer was 19.53 ± 2.23. Then, it peaked at 24.80 ± 1.08 at 3 months (P < 0.0001), decreased to 21.75 ± 1.11 at 6 months (P < 0.0001), and reached 19.77 ± 2.26 at 12 months (P = 0.047). There was no significant correlation between the visual acuity changes and the total corneal densitometry.
Conclusion: The TE-CXL group showed a better and earlier recovery from the haze, while the A-CXL group showed a delay in recovering and persistent increased corneal densitometry, mainly in the anterior 120 μ.

Keywords: Accelerated, Corneal Haze, Cross-linking, Densitometry, Keratoconus


How to cite this article:
Badawi AE. Corneal haze and densitometry in keratoconus after collagen cross-linking by three different protocols. J Curr Ophthalmol 2021;33:422-30

How to cite this URL:
Badawi AE. Corneal haze and densitometry in keratoconus after collagen cross-linking by three different protocols. J Curr Ophthalmol [serial online] 2021 [cited 2022 Aug 14];33:422-30. Available from: http://www.jcurrophthalmol.org/text.asp?2021/33/4/422/335023




  Introduction Top


Since corneal collagen cross-linking (CXL) was firstly introduced at the Technical University of Dresden in Germany in the 1990s,[1] the keratoconus (KC) treatment option was radically evolved, aiming to slow disease progression and reinforce the corneal stromal tissue, thus reducing the need for keratoplasty.

Notwithstanding the favorable clinical and refractive results,[2],[3],[4] CXL–like other surgical interventions–has some early and late complications that should be taken into consideration.[5]

Following the conventional protocol, post-CXL corneal haze is a common clinical consequence with an incidence of up to 90%.[6] The exact causes of corneal haze after CXL are not obviously understood. Keratocytes apoptosis activated keratocytes, corneal lamellae remodeling,[7],[8] extracellular matrix, associated honeycomb-lacunar edema, new CXL bonds,[9] and epithelial debridement[10],[11] may robustly contribute to the post-CXL haze. The standard Dresden protocol[12] was evolved and modified in favor of shortening the overall time of the procedure or decreasing the related complications.[13],[14] The current research has been designed to evaluate the post-CXL corneal haze subjectively and objectively and compare the effects of the three different CXL protocols (standard, accelerated, and trans-epithelial [TE]) on postoperative corneal transparency.


  Methods Top


The study was designed as a retrospective cohort study. The local Ethics Committee of Faculty of Medicine, Mansoura University approved this study (IRB: 17.02.17) which was performed in accordance with the principles of the Declaration of Helsinki. The surgical procedures in all groups were conducted by the author after the patients signed a preoperative, written consent form.

Collected data were selected from the medical records of adult patients with previously documented progressive KC who underwent CXL as a primary sole surgical intervention. The inclusion criteria included 18- to 30-year-old patients with progressive KC Grade I to III, according to the Amsler–Krumeich grading system, and a preoperative corneal thickness of 400 μm or more at the thinnest location and completed an 1-year follow-up. Patients with any previous ocular surgeries, current ocular diseases, such as vitreoretinal disorders, glaucoma, or cataract, or additional surgical intervention, such as intrastromal corneal rings, photorefractive keratectomy, or refractive procedures, were excluded from the study. KC progression was documented using two or more of the following criteria: an increase of 1.0 diopter (D) in K max readings, an increase of 1.0 D in the posterior K readings, an increase of 0.50 D or more in spherical equivalent, and a decrease in more than 2% in the pachymetry in the previous 6 months. The patients were retrospectively enrolled and listed according to three different corneal CXL strategies. Group I were subjected to standard CXL (S-CXL), Group II were subjected to TE-CXL, and Group III were subjected to accelerated CXL (A-CXL).

In Group I, the S-CXL group, the central 8 mm corneal epithelium was scraped with a surgical blade. One drop of the Riboflavin solution 0.1% (Ricrolin, Sooft, Italy) was applied to the corneal surface every 3 min for half an hour. This is followed by 30-min exposure to ultraviolet A (UVA) irradiation (370 nm/3 mW/cm2) using a UVA system (CBM X-linker, Italy). Throughout the irradiation time, the riboflavin solution was administered to the corneal surface at 2-min intervals. Finally, the ocular surface was washed well by a balanced salt solution. Then, a therapeutic contact lens was placed on the cornea until total re-epithelization.

In Group II, the TE-CXL protocol was applied. Two drops of Riboflavin (Ricrolin TE 0.1%, Sooft, Italy) were applied to the intact corneal epithelium every 5 min for 30 min without any disruption. The rest of the procedure and irradiation technique was carried out as described in Group I.

In Group III, the A-CXL group, Riboflavin (Vibex Rapid 0.1%, Avedro, USA) was applied to the corneal surface every 2 min for 10 min after the mechanical removal of the 8 mm epithelium from the central region. Then, UVA irradiation was performed for 9 min using an irradiation UVA device (CBM Vega 10 mW X-Linker, Firenze, Italy). The sequential steps that followed were the same as in Group I. [Table 1] displays the parameters of UVA irradiation systems and the riboflavin solution used in the three groups.
Table 1: The parameters of ultraviolet A irradiation systems and riboflavin solutions used in the three groups

Click here to view


Antibiotic eye drops (Vigamox; Alcon, Inc., USA) were given to the patients for 7 days until full epithelial healing had occurred. Then, they were replaced by combined topical steroid and antibiotic eye drops (Tobradex; Alcon, Inc., USA) for a further 3 weeks with gradual tapering. Eye lubricant drops (Refresh Tears; Allergan, Inc., USA) were prescribed for 1 month and continued for 3 months in indicated cases.

The corneal haze assessment was carried out by the author using slit-lamp biomicroscopy. Different stages and severity were classified from Grade 0 to Grade 4 as follows:[15]

  • Grade 0: No haze “clear cornea”
  • Grade 0.5: Minimal haze detected only by a meticulous examination with oblique illumination
  • Grade 1: Haze with clear visibility of the full iris details
  • Grade 2: Haze mildly obscuring the iris details
  • Grade 3: Haze moderately hiding the iris and lens details
  • Grade 4: Severe haze with total obstruction of the anterior chamber details.


According to the Scheimpflug principle, corneal optical densitometry was assessed (backward light scatter) by an add-on corneal densitometry program to the customary software of the Pentacam Scheimpflug camera (Oculus Pentacam®; Oculus Optikgerate GmbH, Wetzlar, Germany). All measurements were conducted in a dim room preoperatively and at 3 6, and 12 months after the operation. They were repeated three times in all the patients. The program automatically located the corneal apex and the densitometry map. Then, a 12-mm-diameter area around the apex was displayed. For a more elaborate analysis, the corneal densitometry values were recorded in different corneal layers and for each concentric zone. According to the device software, the values of the corneal densitometry of the whole cornea were displayed for three layers based on the depth, the anterior layer (anterior 120 μm), the posterior layer (posterior 60 μm), and the central layer without a particular thickness (between the anterior and the posterior layers). In addition, the total densitometry values of a 12-mm diameter area were exhibited in four concentric zones: 0-2 mm, 2-6 mm, 6-10 mm, and 10-12 mm. The first three zones were included in the comparison between the groups [Figure 1].
Figure 1: A demonstration of corneal densitometry measurement in a keratoconus patient before the accelerated cross-linking (a) and after 6 months (b)

Click here to view


Statistical analysis

Statistical analysis was performed using IBM Statistical Package for Social Science (SPSS) software (version 20.00, SPSS, Chicago, IL, USA). The normality of variables' distribution was tested using the Shapiro–Wilks test. A repeated-measures analysis of variance (ANOVA) test was used to analyze and compare the postoperative changes in each group at each time point. In addition, a one-way ANOVA test was used to compare the three groups. Furthermore, a linear regression test was used for correlation analyses between the best corrected visual acuity at each time point and the changes in the corneal haze after each protocol. A P < 0.05 was statistically significant.


  Results Top


Among 216 eyes subjected to CXL (94 in the S-CXL group, 50 in the TE-CXL group, and 72 in the A-CXL group), only 104 eyes fulfilled the inclusion criteria and were included in this study. The total included eyes were 104 of 53 patients with progressive KC. Group I included 41 eyes (21 patients), Group II included 29 eyes (15 patients), and Group III included 34 eyes (17 patients).

At the end of 1 year, the accelerated epithelial-off CXL (10 mW/cm2 for 9 min) appeared to have similar effects to the S-CXL on slowing the progression of KC in adults while the TE-CXL showed visual and topographic regression at 12 months to reach the preoperative values after an initial improvement. Apart from the corneal haze and densitometry, [Table 2] summarizes the statistical analysis of all estimated pre and postoperative parameters in the three groups.
Table 2: The mean pre and post cross-linking parameters in all studied groups and its P values

Click here to view


The analysis of the mean values of subjective corneal haze demonstrated significant differences between the three groups (P < 0.0001) with better results in favor of Group II. [Figure 2] shows the changes in subjective corneal haze in the three groups at the follow-up time points.
Figure 2: Means of subjective corneal haze in the three groups at the follow-up time points

Click here to view


During the early postoperative period (1–3 months) in Group I, the majority of the eyes (65.7%) had Grade 2 corneal haze, 20% of the eyes had Grade 3, and 14.3% had Grade 1. The corneal haze severity continued to improve until it reached somewhere near the preoperative value at 6 months with the majority having grade (0–0.5).

Three months postoperative in Group II, the majority of the eyes (74.1%) had Grade 1, (18.5%) of the eyes had Grade 2, and (7.4%) had Grade 3. By the 6th month, nearly all eyes showed clear corneas subjectively.

At the end of 6 months in Group III, about 67.5% of the eyes had Grade 2, 20% had Grade 1, and only 12.5% had Grade 3. At the end of 1 year, 82.5% of the eyes had Grade 0.5, 10% had Grade 1, and only 7.5% had Grade 3. [Figure 3] shows the slit-lamp biomicroscopic image of corneal haze in different two cases treated by the S-CXL and the A-CXL protocols.
Figure 3: Slit-lamp biomicroscopic photographs of corneal haze after collagen cross-linking (CXL). Corneal haze Grade 1 in a patient treated with the standard corneal collagen CXL (Group I) 3 months after surgery (a). Corneal haze Grade 2 in a patient treated with the accelerated corneal collagen CXL (Group III) 3 months after surgery (b)

Click here to view


In Group I, the mean densitometry values of all layers and zones were statistically significantly elevated at the end of 3 months compared with its baseline values (P < 0.05). Then, they decreased at 6 months with statistically significant changes between the values at 6 months and baseline in the anterior layer and zones 0–2 mm and 2–6 mm of the central layer with (P = 0.006 and P = 0.001, respectively). The preoperative mean densitometry in the anterior stroma was 19.42 ± 1.81, peaked at 23.12 ± 1.21 at 3 months (P < 0.0001), and reached 19.82 ± 1.19 at 6 months (P = 0.007). At the end of 1 year, the values of corneal densitometry in all layers approached the baseline values with insignificant differences (P > 0.05).

In Group II, there was a significant peak in all the values of all layers at 3 months that were decreased at 6 months with statistically insignificant differences from the baseline values (P > 0.05). Throughout 12 months, the values remained insignificantly changed from the baseline values (P > 0. 05). The preoperative mean densitometry in the anterior stroma was 19.41 ± 1.21, peaked at 19.72 ± 1.12 at 3 months (P = 0.02), decreased to 19.04 ± 1.18 at 6 months (P = 0.052), and reached 19.13 ± 1.37 at 12 months (P = 0.84).

In Group III, the mean corneal densitometry values remained significantly higher than the baseline values of the anterior and central layers mainly zones (0–2 mm) and (2–6 mm) with (P < 0.05) although they were significantly decreased at six and 12 months compared with their measurements at 3 months. The preoperative mean densitometry in the anterior stroma was 19.53 ± 2.23, peaked at 24.80 ± 1.08 at 3 months (P < 0.0001), decreased to 21.75 ± 1.11 at 6 months (P < 0.0001), and reached 19.77 ± 2.26 at 12 months (P = 0.047).

[Figure 4] and [Figure 5] show the changes in the corneal densitometry values of the different corneal layers and zones in the three groups at the follow-up time points.
Figure 4: Corneal densitometry in the different corneal layers in the three groups at the follow-up time points

Click here to view
Figure 5: Corneal densitometry of the three concentric zones (anterior and central layers) at different time points

Click here to view


Despite an insignificant difference in corneal densitometry preoperatively, the densitometry in all stromal layers reached a peak by the end of 3 months with a significant difference between the three groups (P < 0.0001). By the end of 6 months, the densitometry values were decreased with a statistically significant difference between the S-CXL and the TE-CXL groups in the anterior layer; the mean densitometry values of the total anterior layer were 19.82 ± 1.19 in the S-CXL group and 19.04 ± 1.18 in the TE-CXL group (P = 0.049) while the total densitometry values in the central and the posterior layers showed an insignificant difference (P = 0.051 and 0.055, respectively). However, the densitometry readings at 6 months in the A-CXL group revealed significantly higher values than the other two groups in all layers mainly zones (0–2 and 2–6 mm) with P < 0.0001 [Figure 4] and [Figure 5].

By the end of 1-year, corneal densitometry readings decreased to roughly reach the preoperative values in Groups I and II with insignificant differences between them. However, the significant difference between Groups III and the other two groups was still present, predominantly in the anterior and the middle layers (0–2 and 2–6 zones). [Table 3] demonstrates in detail the mean corneal densitometry values postoperatively at 12 months in all studied groups, their differences from the preoperative readings, and the P values.
Table 3: The mean postoperative corneal densitometry at 12 months in all studied groups and P values

Click here to view


For more details, [Supplementary Table 1] shows the mean preoperative corneal densitometry in all studied groups, [Supplementary Table 2] and [Supplementary Table 3] show the mean postoperative corneal densitometry at 3 months and 6 months, respectively.



Corneal densitometry showed an insignificant correlation of the BCVA postoperatively at each time point in the three groups (P > 0 05). After 1 year, the correlation coefficients (r) were 0.021 in the S-CXL group, 0.059 in the TE-CXL group, and 0.072 in the A-CXL group. There were no serious intra or postoperative complications in any group.


  Discussion Top


In the current study, the mean values of corneal densitometry were peaked at the end of 3 months postoperatively with significant differences among the three groups. This is in agreement with a previously reported study by Alzahrani et al.[16]

Besides, there was a significant variance in the improvement of corneal haze post-CXL treatment among the three groups. The A-CXL group started to improve between 6 and 12 months, the corneal densitometry values decreased with time, drawing near, but not reaching, the baseline values. The corneal densitometry in the S-CXL group started to improve at 6 months and draw near the baseline values at 12 months while the TE-CXL group recorded a better and earlier recovery from the haze than the other two groups, it started to improve between three and 6 months and reached the preoperative values nearly by the end of 6 months. These results were consistent with the previous studies of corneal densitometry.[17],[18],[19] However, Greenstein et al.[6] stated that a significant decrease in densitometry and slit-lamp haze was noted between the values at 6 and 12 months after the S-CXL protocol, but they did not reach the baseline values at 1 year. Shen et al.[20] found that the densitometry values were significantly decreased in KC eyes at 1 year after the A-CXL protocol, but they were still higher than the values in myopic individuals. Moreover, Lai et al.[21] reported that, after TE-CXL, corneal haze peaked at 1 month, plateaued between 1 and 3 months, and reached baseline values between 3 and 12 months.

Since the same parameters of the UV rays were used in Groups I and II, in the author's estimation, this might be owing to two factors. The first factor is the role of the epithelium in the corneal haze path.[22] It is known that corneal epithelial injury is one of the main factors in the complex corneal wound healing process.[23] The prominent initial alteration after the epithelial injury is the keratocyte apoptosis, along with associated inflammatory reactions in the anterior corneal stroma. This triggers the production of mature myofibroblasts and promotes an unorganized extracellular matrix secretion.[24] When these changes become settled in the anterior stroma, corneal haze becomes evident. Once myofibroblasts apoptosis is established and keratocytes occupy the anterior corneal stroma, the corneal haze recedes with time.[25] This may explain that the lesser effect the TE-CXL has, the lesser keratocyte damage there is. The second factor is riboflavin absorption behavior. Hypo-osmolar riboflavin appears to be highly unsteady with a short break-up time (90 s) compared with isotonic riboflavin (22 min). Therefore, it has limited absorption and less efficiency of penetration.[25] On the other hand, Razmjoo et al. stated that intact corneal epithelium had an insignificant role in decreasing corneal haze after the CXL protocol.[10]

In the present study, a considerable number of eyes in Group III required a longer time than the other two groups to get back to the baseline level of corneal haze. Some investigators had theorized that the equal efficiency of A-CXL compared to the conventional CXL[26],[27] while others theorized that the conventional CXL protocol was more effective with a deeper demarcation mark.[28],[29] The inverse correlation between UV ray intensity and CXL reaction depth might be in favor of the standard protocol.[30] This controversy could require further evaluation.

Since the same photon numbers interact with the collagen fibrils in different CXL protocols, it is conceivable that the current results (higher and persistent corneal density in the A-CXL group) may be an outcome of the difference in soaking duration and the concentrations of different riboflavin solutions.[31],[32] Corneal stromal saturation and absorption traits of different riboflavin solutions were evaluated by several studies.[33],[34] It was reported that the riboflavin concentration during the epithelial-off CXL is about seven-fold higher in the anterior stroma compared with its concentration in the epithelial-on CXL, which is well correlated with the density of keratocytes apoptosis and the depth of the demarcation line.[35] Despite the constant homogeneous consumption of riboflavin in both the standard and the accelerated protocols, the consumption of the riboflavin solution during the S-CXL irradiation of the corneal stroma was stated to be 20% more than that achieved during the A-CXL one (consumption of riboflavin was 87% ± 2% in the S-CXL instead of 67% ±3% in the A-CXL).[36] This might explain that there is a higher concentration of the riboflavin solution (VibeX Rapid 0.1%, Avedro, USA) in the anterior corneal stroma of the A-CXL group than the riboflavin solution (Ricrolin 0.1%, Sooft, Italy) in the conventional CXL. Further investigations should be conducted on this topic to clarify the relation between the different riboflavin concentrations and the CXL efficiency as well as the corneal healing process.

Another explanation of the current results is that the substantial role of oxygen in the CXL photochemical reaction improved the polymerization outcomes. The S-CXL protocol was theorized to permit stable oxygen consumption and further re-diffusion out of the corneal tissue.[37] Theoretically, the A-CXL protocol might lead to inadequate time for oxygen diffusion. High-irradiance CXL was suggested to be the cause of reducing the oxygen diffusion capacity and increasing its consumption.[38] Currently, continuous A-CXL was used rather than pulsed CXL. Continuous epithelium-off A-CXL appears to cause more corneal haze than pulsed epithelium-off A-CXL and pulsed TE-CXL, especially in the early postoperative period.[39]

It was hypothesized that the pulsed CXL light was better in achieving sufficient oxygen diffusion required for the CXL reaction.[40],[41] The reduction of the oxygen level below the certain perilous threshold will cause the tissue to return to aerobic metabolism42 and lose the corneal tissue transparency as one of the clinical consequences that ensued from the aerobic metabolic process.[43] Hypoxia also was reported as a factor that affects cell migration and corneal wound healing.[44] This might explain the delay in recovering from the haze in Group III. This is another interesting topic for further research to reveal whether or not hypoxia associated with high-irradiance CXL affects the healing process.

It is worth mentioning here that despite the difference in the corneal haze and density between the groups, it did not affect the postoperative BCVA. There were insignificant correlations between the corneal haze or density changes and the BCVA changes at each time point. Our results supported earlier reports of a poor correlation between haze and visual acuity (VA).[6],[45] However, the study was looking at the changes in corneal haze and density as a reflection of the wound healing process after different CXL protocols rather than its effect on VA. Similarly, Bouheraoua et al.[45] found no remarkable correlations between the CXL demarcation line depth after CXL and the BCVA changes at 6 months. On the other hand, Mathews et al. stated a significant correlation between the change in corneal densitometry and the decrease in higher-order aberrations at 6 months post-CXL.[46] Larger, prospective, randomized studies may be required to study this topic.

The study included some limitations such as small sample size, retrospective design, lack of histopathological study, and absence of demarcation-line depth measures.

In conclusion, there were significant variations in corneal densitometry in response to the different CXL protocols. The A-CXL group showed a delay in recovering from the haze with persistent increased corneal densitometry, mainly in the anterior 120 μ after 1 year from the surgery. There were no significant correlations between the changes in the BCVA and the changes in the corneal densitometry over time. At the end of 1 year, the accelerated epithelial-off CXL appeared to have similar effects, such as the S-CXL protocol, in slowing the progression of KC in adults. However, the TE-CXL showed visual and topographic regression that warrants more future studies with longer follow-up.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
  References Top

1.
Raiskup-Wolf F, Hoyer A, Spoerl E, Pillunat LE. Collagen crosslinking with riboflavin and ultraviolet-A light in keratoconus: Long-term results. J Cataract Refract Surg 2008;34:796-801.  Back to cited text no. 1
    
2.
Vinciguerra P, Albè E, Trazza S, Rosetta P, Vinciguerra R, Seiler T, et al. Refractive, topographic, tomographic, and aberrometric analysis of keratoconic eyes undergoing corneal cross-linking. Ophthalmology 2009;116:369-78.  Back to cited text no. 2
    
3.
Grewal DS, Brar GS, Jain R, Sood V, Singla M, Grewal SP. Corneal collagen crosslinking using riboflavin and ultraviolet-A light for keratoconus: One-year analysis using Scheimpflug imaging. J Cataract Refract Surg 2009;35:425-32.  Back to cited text no. 3
    
4.
Caporossi A, Mazzotta C, Baiocchi S, Caporossi T. Long-term results of riboflavin ultraviolet a corneal collagen cross-linking for keratoconus in Italy: The Siena eye cross study. Am J Ophthalmol 2010;149:585-93.  Back to cited text no. 4
    
5.
Dhawan S, Rao K, Natrajan S. Complications of corneal collagen cross-linking. J Ophthalmol 2011;2011:869015.  Back to cited text no. 5
    
6.
Greenstein SA, Fry KL, Bhatt J, Hersh PS. Natural history of corneal haze after collagen crosslinking for keratoconus and corneal ectasia: Scheimpflug and biomicroscopic analysis. J Cataract Refract Surg 2010;36:2105-14.  Back to cited text no. 6
    
7.
Wollensak G, Herbst H. Significance of the lacunar hydration pattern after corneal cross linking. Cornea 2010;29:899-903.  Back to cited text no. 7
    
8.
Mencucci R, Marini M, Paladini I, Sarchielli E, Sgambati E, Menchini U, et al. Effects of riboflavin/UVA corneal cross-linking on keratocytes and collagen fibres in human cornea. Clin Exp Ophthalmol 2010;38:49-56.  Back to cited text no. 8
    
9.
Mazzotta C, Balestrazzi A, Baiocchi S, Traversi C, Caporossi A. Stromal haze after combined riboflavin-UVA corneal collagen cross-linking in keratoconus: In vivo confocal microscopic evaluation. Clin Exp Ophthalmol 2007;35:580-2.  Back to cited text no. 9
    
10.
Razmjoo H, Rahimi B, Kharraji M, Koosha N, Peyman A. Corneal haze and visual outcome after collagen crosslinking for keratoconus: A comparison between total epithelium off and partial epithelial removal methods. Adv Biomed Res 2014;3:221.  Back to cited text no. 10
    
11.
Shalchi Z, Wang X, Nanavaty MA. Safety and efficacy of epithelium removal and transepithelial corneal collagen crosslinking for keratoconus. Eye (Lond) 2015;29:15-29.  Back to cited text no. 11
    
12.
Wollensak G, Spoerl E, Seiler T. Riboflavin/ultraviolet-a-induced collagen crosslinking for the treatment of keratoconus. Am J Ophthalmol 2003;135:620-7.  Back to cited text no. 12
    
13.
Hashemi H, Fotouhi A, Miraftab M, Bahrmandy H, Seyedian MA, Amanzadeh K, et al. Short-term comparison of accelerated and standard methods of corneal collagen crosslinking. J Cataract Refract Surg 2015;41:533-40.  Back to cited text no. 13
    
14.
De Bernardo M, Capasso L, Tortori A, Lanza M, Caliendo L, Rosa N. Trans epithelial corneal collagen crosslinking for progressive keratoconus: 6 months follow up. Cont Lens Anterior Eye 2014;37:438-41.  Back to cited text no. 14
    
15.
Fantes FE, Hanna KD, Waring GO 3rd, Pouliquen Y, Thompson KP, Savoldelli M. Wound healing after excimer laser keratomileusis (photorefractive keratectomy) in monkeys. Arch Ophthalmol 1990;108:665-75.  Back to cited text no. 15
    
16.
Alzahrani K, Mofty H, Lin EY, Carley F, Brahma A, Morley D, et al. Corneal imaging and densitometry measurements in juvenile and adult keratoconus patients to evaluate disease progression and treatment effects after corneal cross-linking. Clin Optom (Auckl) 2019;11:173-80.  Back to cited text no. 16
    
17.
Koller T, Mrochen M, Seiler T. Complication and failure rates after corneal crosslinking. J Cataract Refract Surg 2009;35:1358-62.  Back to cited text no. 17
    
18.
Gutiérrez R, Lopez I, Villa-Collar C, González-Méijome JM. Corneal transparency after cross-linking for keratoconus: 1-year follow-up. J Refract Surg 2012;28:781-6.  Back to cited text no. 18
    
19.
Kim BZ, Jordan CA, McGhee CN, Patel DV. Natural history of corneal haze after corneal collagen crosslinking in keratoconus using Scheimpflug analysis. J Cataract Refract Surg 2016;42:1053-9.  Back to cited text no. 19
    
20.
Shen Y, Jian W, Sun L, Li M, Han T, Son J, et al. One-year follow-up of changes in corneal densitometry after accelerated (45 mW/cm2) transepithelial corneal collagen cross-linking for keratoconus: A retrospective study. Cornea 2016;35:1434-40.  Back to cited text no. 20
    
21.
Lai M, Greenstein SA, Gelles JD, Hersh PS. Corneal haze after transepithelial collagen cross-linking for keratoconus: A scheimpflug densitometry analysis. Cornea 2020;39:1117-21.  Back to cited text no. 21
    
22.
Nakamura K, Kurosaka D, Bissen-Miyajima H, Tsubota K. Intact corneal epithelium is essential for the prevention of stromal haze after laser assisted in situ keratomileusis. Br J Ophthalmol 2001;85:209-13.  Back to cited text no. 22
    
23.
Netto MV, Mohan RR, Sinha S, Sharma A, Dupps W, Wilson SE. Stromal haze, myofibroblasts, and surface irregularity after PRK. Exp Eye Res 2006;82:788-97.  Back to cited text no. 23
    
24.
Torricelli AA, Santhanam A, Wu J, Singh V, Wilson SE. The corneal fibrosis response to epithelial-stromal injury. Exp Eye Res 2016;142:110-8.  Back to cited text no. 24
    
25.
Torricelli AA, Singh V, Agrawal V, Santhiago MR, Wilson SE. Transmission electron microscopy analysis of epithelial basement membrane repair in rabbit corneas with haze. Invest Ophthalmol Vis Sci 2013;54:4026-33.  Back to cited text no. 25
    
26.
Kanellopoulos AJ. Long term results of a prospective randomized bilateral eye comparison trial of higher fluence, shorter duration ultraviolet A radiation, and riboflavin collagen crosslinking for progressive keratoconus. Clin Ophthalmol 2012;6:97-101.  Back to cited text no. 26
    
27.
Cınar Y, Cingü AK, Türkcü FM, Çınar T, Yüksel H, Özkurt ZG, et al. Comparison of accelerated and conventional corneal collagen cross-linking for progressive keratoconus. Cutan Ocul Toxicol 2014;33:218-22.  Back to cited text no. 27
    
28.
Chow VW, Chan TC, Yu M, Wong VW, Jhanji V. One-year outcomes of conventional and accelerated collagen crosslinking in progressive keratoconus. Sci Rep 2015;5:14425.  Back to cited text no. 28
    
29.
Spadea L, Tonti E, Vingolo EM. Corneal stromal demarcation line after collagen cross-linking in corneal ectatic diseases: A review of the literature. Clin Ophthalmol 2016;10:1803-10.  Back to cited text no. 29
    
30.
Ng AL, Chan TC, Cheng AC. Conventional versus accelerated corneal collagen cross-linking in the treatment of keratoconus. Clin Exp Ophthalmol 2016;44:8-14.  Back to cited text no. 30
    
31.
Cingü AK, Sogutlu-Sari E, Cınar Y, Sahin M, Türkçü FM, Yüksel H, et al. Transient corneal endothelial changes following accelerated collagen cross-linking for the treatment of progressive keratoconus. Cutan Ocul Toxicol 2014;33:127-31.  Back to cited text no. 31
    
32.
Spoerl E, Mrochen M, Sliney D, Trokel S, Seiler T. Safety of UVA-riboflavin cross-linking of the cornea. Cornea 2007;26:385-9.  Back to cited text no. 32
    
33.
Søndergaard AP, Hjortdal J, Breitenbach T, Ivarsen A. Corneal distribution of riboflavin prior to collagen cross-linking. Curr Eye Res 2010;35:116-21.  Back to cited text no. 33
    
34.
Iseli HP, Popp M, Seiler T, Spoerl E, Mrochen M. Laboratory measurement of the absorption coefficient of riboflavin for ultraviolet light (365 nm). J Refract Surg 2011;27:195-201.  Back to cited text no. 34
    
35.
Franch A, Birattari F, Dal Mas G, Lužnik Z, Parekh M, Ferrari S, et al. Evaluation of intrastromal riboflavin concentration in human corneas after three corneal cross-linking imbibition procedures: A pilot study. J Ophthalmol 2015;2015:794256.  Back to cited text no. 35
    
36.
Lombardo G, Micali NL, Villari V, Serrao S, Lombardo M. All-optical method to assess stromal concentration of riboflavin in conventional and accelerated UV-A irradiation of the human cornea. Invest Ophthalmol Vis Sci 2016;57:476-83.  Back to cited text no. 36
    
37.
Kamaev P, Friedman MD, Sherr E, Muller D. Photochemical kinetics of corneal cross-linking with riboflavin. Invest Ophthalmol Vis Sci 2012;53:2360-7.  Back to cited text no. 37
    
38.
Caporossi A, Mazzotta C, Paradiso AL, Baiocchi S, Marigliani D, Caporossi T. Transepithelial corneal collagen crosslinking for progressive keratoconus: 24-month clinical results. J Cataract Refract Surg 2013;39:1157-63.  Back to cited text no. 38
    
39.
Ziaei M, Gokul A, Vellara H, Patel D, McGhee CN. Prospective two year study of changes in corneal density following transepithelial pulsed, epithelium-off continuous and epithelium-off pulsed, corneal crosslinking for keratoconus. Cont Lens Anterior Eye 2020;43:458-64.  Back to cited text no. 39
    
40.
Mazzotta C, Traversi C, Paradiso AL, Latronico ME, Rechichi M. Pulsed light accelerated crosslinking versus continuous light accelerated crosslinking: One-year results. J Ophthalmol 2014;2014:604731.  Back to cited text no. 40
    
41.
Akkaya Turhan S, Toker E. Changes in corneal density after accelerated corneal collagen cross-linking with different irradiation intensities and energy exposures: 1-year follow-up. Cornea 2017;36:1331-5.  Back to cited text no. 41
    
42.
National Research Council (US) Working Group on Contact Lens Use Under Adverse Conditions. In: Ebert Flattau P, editor. Considerations in Contact Lens Use Under Adverse Conditions: Proceedings of a Symposium. Washington (DC): National Academies Press (US); 1991. Hypoxia. Available from: https://www.ncbi.nlm.nih.gov/books/NBK234101/. [Last accessed on 2021 Jul 05].  Back to cited text no. 42
    
43.
Leung BK, Bonanno JA, Radke CJ. Oxygen-deficient metabolism and corneal edema. Prog Retin Eye Res 2011;30:471-92.  Back to cited text no. 43
    
44.
Onochie O, Trinkaus-Randall V. Hypoxia affects cell migration and wound healing by altering basal lamina proteins and focal adhesion turnover. FASEB J 2015;29:719-4.  Back to cited text no. 44
    
45.
Bouheraoua N, Jouve L, El Sanharawi M, Sandali O, Temstet C, Loriaut P, et al. Optical coherence tomography and confocal microscopy following three different protocols of corneal collagen-crosslinking in keratoconus. Invest Ophthalmol Vis Sci 2014;55:7601-9.  Back to cited text no. 45
    
46.
Mathews PM, De Rojas JO, Rapuano PB, Zemsky CJ, Florakis GJ, Trokel SL, et al. Correlation of Scheimpflug densitometry changes with clinical outcomes after corneal crosslinking. J Cataract Refract Surg 2018;44:993-1002.  Back to cited text no. 46
    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5]
 
 
    Tables

  [Table 1], [Table 2], [Table 3]



 

Top
 
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
Abstract
Introduction
Methods
Results
Discussion
References
Article Figures
Article Tables

 Article Access Statistics
    Viewed1077    
    Printed32    
    Emailed0    
    PDF Downloaded111    
    Comments [Add]    

Recommend this journal


[TAG2]
[TAG3]
[TAG4]