ORIGINAL ARTICLE
Year : 2022  |  Volume : 34  |  Issue : 1  |  Page : 44-49

Differences and limits of agreement among pentacam, Corvis-ST, and IOL-Master 700 optical biometric devices regarding central corneal thickness measurements


1 Department of Optometry, School of Rehabilitation Sciences, Iran University of Medical Sciences, Tehran, Iran
2 Department of Optometry, Iran University of Medical Sciences, Tehran, Iran
3 Department of Basic Sciences, School of Nursing and Midwifery, Shahid Beheshti University of Medical Sciences, Tehran, Iran
4 Internal Medical Clinic, Bank Melli Hospital, Tehran, Iran

Correspondence Address:
Abbas Riazi
School of Rehabilitation Sciences, Iran University of Medical Sciences, Maddadkaran St., Shahid Nazari St., Mother Square, Mirdamad Blvd, P. O. Box: 15875-4391, Tehran
Iran
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/joco.joco_96_21

Rights and Permissions

Purpose: To investigate the differences and limits of agreement in measuring corneal thickness using Pentacam, Corvis, and intraocular lens (IOL)-Master 700 devices. Methods: This study was conducted on 37 right eyes of 21 males and 16 females (n = 37) with a mean age of 52.11 ± 6.30 years. The central corneal thickness was measured using three optical biometric devices, including Pentacam, Corvis, and IOL-Master 700. The inclusion criteria were normal eyes without any ophthalmological abnormalities, history of ocular pathology, or ocular surgery. The data obtained from these three devices were compared two by two. The correlation and agreement limits among them were analyzed using statistical techniques. Results: The mean standard deviation differences between Pentacam and Corvis, Pentacam and IOL-Master 700, as well as Corvis and IOL-Master 700 regarding the corneal thickness measurement, were 22.13 ± 8.05, 7.91 ± 8.02, and 14.21 ± 9.85 μm, respectively, which were statistically significant (P < 0.0001). Based on the investigation of the limits of agreement according to the Bland Altman method, the corresponding values between Pentacam and Corvis, Pentacam and IOL-Master 700, and Corvis and IOL-Master 700 were -16.2 to +15.4, -15.8 to +16.3, and -20.1 to +20.0 μm, respectively. Furthermore, the correlation coefficients of the measurements obtained by Pentacam and Corvis, Pentacam and IOL-Master 700, as well as Corvis and IOL-Master 700 were determined 0.957, 0.964, and 0.948, respectively (P < 0.0001). Conclusion: The results from this study indicate that the interchangeable use of these three devices is not appropriate due to statistically significant differences and broad limits of agreement among the three devices, especially between Corvis and IOL-Master 700.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed248    
    Printed6    
    Emailed0    
    PDF Downloaded28    
    Comments [Add]    

Recommend this journal