ORIGINAL ARTICLE |
|
Year : 2022 | Volume
: 34
| Issue : 1 | Page : 67-73 |
|
Assessing accuracy of okulix ray-tracing software in calculating intraocular lens power in the long cataractous eyes
Alireza Jamali1, Mahmoud Jabbarvand2, Amir Asharlous1, Asgar Doostdar1, Soodeh Bordbar3, Mehdi Khabazkhoob4
1 Rehabilitation Research Center, Department of Optometry, School of Rehabilitation Sciences, Iran University of Medical Sciences, Tehran, Iran 2 Eye Research Center, Farabi Eye Hospital, Tehran University of Medical Sciences, Tehran, Iran 3 Department of English, Iran University of Medical Sciences, Tehran, Iran 4 Department of Psychiatric Nursing and Management, School of Nursing and Midwifery, Shahid Beheshti University of Medical Sciences, Tehran, Iran
Correspondence Address:
Asgar Doostdar Rehabilitation Research Center, Department of Optometry, School of Rehabilitation Sciences, Iran University of Medical Sciences, Tehran Iran
 Source of Support: None, Conflict of Interest: None
DOI: 10.4103/joco.joco_136_20
|
|
Purpose: To investigate the accuracy of Okulix ray-tracing software in calculating intraocular lens (IOL) power in the long cataractous eyes and comparing the results with those obtained from Kane, Holladay 1 with optimized constant, SRK/T with optimized constant, Haigis with optimized constant, and Barret Universal 2 formulas.
Methods: The present study evaluates the refractive results of cataract surgery in 85 eyes with axial length > 25 mm and no history of ocular surgery and corneal pathology. IOL power calculation was performed using the Okulix software. The performances of Okulix software in comparison with the five other formulas were evaluated by predicted error, mean absolute error, and mean numerical error 6 months after surgery.
Results: The mean calculated IOL power by the Okulix software was +13.48 ± 4.19 diopter (D). The mean of the 6-month postoperative sphere and spherical equivalent were +0.18 ± 0.63 and -0.34 ± 0.78 D, respectively. Also, the 6-month spherical equivalent in 56.6% and 80% of eyes were within ±0.05 and ±1.00 D, respectively. The predicted error (P < 0.001) and the mean numerical error (P < 0.001) were different between the six studied methods; however, we were not able to find any significant differences in the mean absolute error among six studied methods (P: 0.211).
Conclusion: The present study showed acceptable performance of the Okulix software in IOL power calculation in long eyes in comparison with the other five methods based on the postoperative refractive error, calculated mean absolute error, and mean numerical error.
|
|
|
|
[FULL TEXT] [PDF]* |
|
 |
|